TH8145 Halogen acid gas release measuring device
Overview:
It is used to determine the total amount of halogen acid gas in low-halogen and halogen-free flame-retardant cables and to measure the acidity of the gas by measuring the PH value and conductivity.
Comply with standards: GB/T17650.1-1998, GB/T17650.2-1998 and IEC60754-1:1994, IEC60754-2:1991 standards; at the same time meet the GB/T19666-2005 "Flame-retardant and fire-resistant wires and cables" standard Table 6 regulations and GA306.1-2007 Article 6.3.4, GA306.2-2007 Article 6.3.4, UL2556-2007 Article 9.10 and 9.11 requirements.
Main technical parameters:
1. Tubular furnace: The effective length and diameter of the heating chamber are 600× ф42mm, equipped with an adjustable electric heating system. Heating power: 220V /10KW
2. Combustion tube:
1. Material: Refractory pipe made of silica, resistant to corrosive gas.
2. Specifications: The inner diameter is 40mm, the length extending from the inlet side of the tube furnace is 60~200mm, and the length extending from the outlet side is 60~100mm, leaving a gap for thermal expansion.
3. Burning boat for measuring the release of halogen acid gas:
1. Material: Use fused quartz tube.
2. Specifications: length 100mm, width 25mm, depth 10mm.
3. Combustion boat feeding method: using bar magnet and platinum wire pulling method.
4. Gas collection: Two selected bottles are used: one of which is equipped with a magnetic stirrer and contains 220 mL of sodium hydroxide solution.
5. Gas supply system:
1. Air flow meter: 20mL~200mL adjustable
2. Compressor (provided by your company)
6. Sampling: 500mg~1000mg
7. Determination of the halogen acid content of the halogen acid gas release measuring device:
1. Analytical balance: accuracy of 0.1mg
2. Digital PH meter: the accuracy is 0.02PH
3. Conductivity meter: Measuring range: (0~1×105)μS/cm.