Product Center
FuyangTH8145 Halogen Acid Gas Release Measurement Device
Category:
FuyangPlastic testing instruments
Product Description
Overview:
It is used to determine the total halogen acid gas content in low-halogen and halogen-free flame-retardant cables, as well as to assess the acidity of the gases by measuring pH and electrical conductivity.
Complies with the standards GB/T 17650.1‑1998, GB/T 17650.2‑1998, IEC 60754‑1:1994, and IEC 60754‑2:1991; and also meets the requirements specified in Table 6 of GB/T 19666‑2005 “General Requirements for Flame‑Retardant and Fire‑Resistant Wires and Cables,” as well as Clauses 6.3.4 of GA 306.1‑2007 and GA 306.2‑2007, and Clauses 9.10 and 9.11 of UL 2556‑2007.
Main technical parameters:
1. Tubular Furnace: The heating chamber has an effective length of 600 mm and a tube diameter of 42 mm, equipped with an adjustable electric heating system. Heating power: 220 V / 10 kW.
II. Combustion Tube:
1. Material: Refractory tube made of silicon dioxide, resistant to corrosive gases.
2. Specifications: Inner diameter 40 mm; the extension length on the inlet side of the tubular furnace is 60–200 mm, and on the outlet side is 60–100 mm, with a gap provided to accommodate thermal expansion.
3. Combustion boat for the halogen acid gas release measurement apparatus:
1. Material: Made of fused quartz tubing.
2. Specifications: Length 100 mm, width 25 mm, depth 10 mm.
3. Combustion boat insertion method: A strip-shaped magnet and platinum wire are used to manipulate the boat.
4. Gas Collection: Two flasks were used; one was equipped with a magnetic stirrer and contained 220 mL of sodium hydroxide solution.
V. Gas Supply System:
1. Air flow meter: adjustable from 20 mL to 200 mL
2. Compressor (provided by your company)
6. Sampling: 500 mg–1000 mg
VII. Determination of Halogen Acid Gas Emissions: Measurement of Halogen Acid Content:
1. Analytical balance: accuracy of 0.1 mg
2. Digital pH meter: accuracy of 0.02 pH units
3. Conductivity meter: Measurement range: (0–1×10⁵) μS/cm.